EZ4006 Chloride Analyser

EZ4006 Chloride Analyser
Product #: EZ4006.XXXXXXXX
Unit Price Contact Us

What's in the box

EZ4006 Chloride Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x mV electrode, 2 x empty 10L Reagent Containers with Fittings (for Cleaning & Buffer Solution) and 1 x empty 10L Reagent Dark Container with Fittings (for AgNO₃ Solution)

Specifications

Alarm: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Ambient temperature: 10 - 30 °C ±4 °C deviation at 5 - 95% relative humidity (non-condensing)
Analogue outputs: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Calibration: N.A.
Certifications: CE compliant / UL certified
Cycle time: 10 - 15 minutes
Demineralised water: For rinsing / dilution
Detection limit: 5 mg/L
Digital outputs: Optional: Modbus (TCP/IP, RS485)
Dimensions (H x W x D): 690 mm x 465 mm x 330 mm
Drain: Atmospheric pressure, vented, min. Ø 32 mm
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Flow rate: 100 - 300 mL/min
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferences: Bromide, sulphide, iodide ions may interfere. Ferricyanide causes high results and must be removed. Chromate and dichromate interfere and should be reduced to chromic state or removed. Ferric iron interferes if present in an amount substantially higher than the amount of chloride. Chromic ion, ferrous ion and phosphate do not interfere. Fats, oil, proteins, surfactants and tar.
Material: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Measurement method: Potentiometric titration with silver nitrate (AgNO₃), conform with standard method APHA 4500-Cl (D)
Measuring range: 25 ‐ 500 mg/L
  25 - 500 mg/L Cl-

Optional:
5 - 50 mg/L Cl-
5 - 125 mg/L Cl-
5 - 250 mg/L Cl-
Internal dispenser dilution (max. factor 100)
Method: Titrimetric
Model: EZ4006
Number of sample streams: 1, 2, 4 or 8
Parameter: Chloride
Power: 100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Power supply: 100 - 240 VAC, 50/60 Hz
Precision: Better than 2% full scale range for standard test solutions
Protection class: Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Reagent requirements: Keep between 10 - 30 °C
Sample pressure: By external overflow vessel
Sample quality: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample temperature: 10 - 30 °C
Validation: Automatic; frequency freely programmable
Warranty: 24 months
Weight: 25 kg
What's included?: EZ4006 Chloride Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x mV electrode, 2 x empty 10L Reagent Containers with Fittings (for Cleaning & Buffer Solution) and 1 x empty 10L Reagent Dark Container with Fittings (for AgNO₃ Solution)