Bühler 4011 Stationary automatic water sampler

Bühler 4011 Stationary automatic water sampler
Product #: BU4011.XX.XXXXX
Unit Price Contact Us
Available

Stainless steel housing (optional: plastic-coated stainless steel)

Strict separation of control and sample sections

Composite and fractionated samples

BUS and GSM communication

Specifications

Ambient temperature: -20 - 43 °C
Bottle filling time: Bottle filling time adjustable 00:01 mm:hh to 999:59 mm:hh
Certifications: CE, Sampling according to ISO 5667-2/3-10
Communication: Optional: Modbus, connection via Profibus DP, PC via Software
Further communication options on request.
Controller: Microprocessor control, Sleep-Mode (<5 mA), power supply 8-16 V, foil keyboard (with keys 0 - 9, ESC, ENT, cursor), graphical display (128 x 64 Pixel), backlit
Cooling system: Self-contained, controlled cooling / heating with 4 settings, no-frost. independent of the programmable controller, temperature in sample compartment: 4 °C (adjustable from 0.0 - 9.9 °C)
Data storage: 3000 entries, non-volatile data memory; storage of sampling and malfunction data like sample extractions, bottle changes, messages, external signals
Dimensions: * with opened top
  1290 (1890*) mm x 690 mm x 645 mm (H x W x D)
Housing material: Double-walled stainless steel (material 1.4301/SS304) / PS / PC (GF10) with 40 mm insulation. Housing separated in sample compartment and control compartment, each with lockable door. Upper door with PMMA window. Protective top made of Styrosun which can be opened for connection and maintenance works.
Optional: 1.4571/SS316Ti; SS304 Epoxy-coated; SS316Ti Epoxy-coated
  Stainless Steel
Inputs: 1 x analogue: 0/4-20 mA
2 x digital (flow, event), Impulse length 60 ms, switching level 7-24 V, max. working resistance 500 Ohm, max. length of signal cable 30 m
Interface: Mini-USB
Keypad: Numeric key pad, 4 arrow navigation keys, enter and back key
Languages user interface: Multi-language, selectable
Logging: Sample history, input signals, bottle changes, alarms in combination with time/date stamp
Manual languages: DE, FR, GB, IT, NL, ES, SE, CZ, DK, RU, PL, RO, BG
Manual sampling: Possible at any time without interrupting the current program run
Non return valve: Available in combination with glass dosing systems
Number of user programs: free to define user programs
Operatin conditions: Pressureless sample taking is a must for all dosing options except Bypass
Overfilling protection: Adjustable from 1 - 999 samples/bottle
Password protection: Programm changes and internal functions can be password protected
Pause mode: Interruption of program run at any time
Power requirements (Voltage): 115/230 V AC
Power supply: Approx. 350 VA (with cooling)
Program protection: Up to 5 years after voltage loss
Programs: 12 freely programmable user programs, with function to link programs
Program start options: Immediately; at a certain time; by an external signal
Program stop options: End of sampling program after one program run; continuous operation or x- runs, and a certain date / time
Relays: Collective malfunction relay as standard, up to 4 relays optionally available
Sample container: Plastic: 1 x 25 L; 1 x 50 L; 2 x 10 L; 2 x 22 L; 4 x 10 L; 4 x 14 L; 4 x 20 L; 4 x 25 L; 12 x 2.9 L; 24 x 1,0 L; 24 x 2.9 L
Glass: 12 x 2.0 L; 24 x 0.9 L; 24 x 2.0 L
Sample distributor: Round distributor for multi-bottle versions (distribution plate used for some bottle configurations)
Sample interval: 1 min to 99 h 59 min in steps of 1 minute
1 to 9999 pulses/sample
Sampler type: Stationary
Sample temperature: 0 - 40°C (32 - 104°F)
Sample transport velocity: >0.5 m/s at suction height up to at least 8 m (at 1013 hPa); pump performance can be adjusted electronically
Sample volume: Sample volume adjustable at standard dosing system 20 - 350 mL (optional Dosing Systems: Bypass, CTVV - constant time, variable volume, 500 mL dosing option)
Sample volume accuracy: Standard Vacuum System: < 2.5 % or ± 3 mL
Sampling mode: Time-related, flow-proportional, constant time/ variable volume (CTVV), event-related, manual sample extraction
Sampling principle: Standard Vacuum System 20 - 350 mL

Option: Vacuum System 20 - 500 mL

Option: Vacuum CTVV flow-proportional 5 - 350 mL

Option: Bypass System 20 - 250 mL

Option: glass dosing vessel instead of plastic
Status messages: Standard: 1 collective malfunction message as relay output
Option: expandable up to 5 relay outputs - freely programmable
Suction Height: Max. 8 m (at 1013 hPa and stagnant medium), optional higher
Suction hose inner diameter: 12 mm
Suction hose length: 7.5 m; max. hose length 30 m
Suction hose material: PVC
Suction hose outer diameter: 14
Type of distributor: Round distributor for multi-bottle versions (distribution plate used for some bottle configurations)
Warranty: 24 months
Weight: Approx. 100 kg with composite container, higher weight when using multi bottle option and/or glass bottles
Wetted materials: PC, PVC, Silicone, PS, PE, EPDM (optional: metering vessel glass Duran50, sinker weight SS304)