EZ1030 Phenol Analyser, 1 stream, Modbus RS485

EZ1030 Phenol Analyser, 1 stream, Modbus RS485
Product #: EZ1030.99001C02
GBP Price: Contact Us

Specifications

Alarm: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Ambient temperature: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Calibration: Automatic, 2-point; frequency freely programmable
Certifications: CE compliant / UL certified
Cycle time: 10 min (dilution + 5 min)
Detection limit: ≤ 5 µg/L
Digital outputs: Modbus RS485
Dimensions (H x W x D): 690 mm x 465 mm x 330 mm
Drain: Atmospheric pressure, vented, min. Ø 64 mm
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Flow rate: 100 - 300 mL/min
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferences: Oxidising agents such as Chlorine, Sulphur compounds like Sulphide. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Material: Hinged part: Thermoform ABS, door: plexiglass
Wall section: Galvanised steel, powder coated
Measurement method: Colorimetric measurement at 505 nm using 4-aminoantipyrine, conform with standard method APHA
Measuring range: 0 - 2 mg/L Phenol

Optional:
0 - 0.2 mg/L
0 - 0.5 mg/L
0 - 1 mg/L
0 - 8 mg/L (with internal dilution)
0 - 16 mg/L (with internal dilution)
0 - 200 mg/L (with internal dilution)
Number of sample streams: 1 stream

Optional:
1 to 8 streams
Output: Modbus RS485

Optional:
Active 4 - 20 mA max. 500 Ohm load, 1 to 8 outputs
RS232, Modbus TCP/IP
Parameter: Phenol
Power: 110 - 220 VAC, 2 A, 50/60Hz
Max. power consumption: 150 VA
Other voltages available on request
Precision: Better than 3% full scale range for standard test solutions
Protection class: Analyser cabinet: IP55 / Panel PC: IP65
Reagent requirements: Keep between 10 - 30 °C
Sample pressure: By external overflow vessel
Sample quality: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample temperature: 10 - 30 °C
Validation: Automatic; frequency freely programmable
Warranty: 2 years
Weight: 25 kg