Polymetron 9611 sc Online Phosphate Analyser

Polymetron 9611 sc Online Phosphate Analyser
Product #: 9611.KTO.PX.XXX
Unit Price Contact Us

Low maintenance

Reduced downtime

User-defined reagents that meet individual needs

90 days of continuous runtime

Easily verified with our lab products - no time wasted by second-guessing

Specifications

Accuracy: LR model: ±4 µg/L or ±4% of reading, whichever is greater;
HR model: ±500 µg/L or ±5% of reading, whichever is greater
(Tested with Hach reagents)
Alarm: Four relays output; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC maximum
Connection: 0.82 to 2.08 mm2 wire (18 to 14 AWG ), 0.82 mm2 stranded (18 AWG) recommended
Application: Pure water / Power
Certifications: CE (EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000)
KC (EN 61326-1: 2006)
C-tick (EN 61326-1: 2006)
cETLus (UL 61010-1: 2012; NEMA 250: 2003; CSA C22.2 No 61010-1: 2012)
Controller compatibility: SC200, SC1000
Detection limit: LR model: 4 µg/L
HR model: 200 µg/L
Dimensions (H x W x D): 804 mm x 452 mm x 360 mm
Enclosure waterproof rating: IP56 / NEMA 4X
Fitting: Sample line and sample bypass drain: 6 mm (¼-in.)
Air purge inlet: 6 mm (¼-in.) Chemical and case drains: 9.5 mm (3/8-in.)
Flow rate: 55 - 300 mL/min
Grab sample: Grab Sample In and Grab Sample Out capability
Light source: Class 1M LED (light emitting diode) with a peak wavelength of 810 nm
Manual languages: English
French
Spanish
B.Portuguese
Chinese
Japanese
Korean
Thai
German
Italian
C.Portuguese
Czech
Danish
Dutch
Polish
Swedish
Finnish
Bulgarian
Hungarian
Romanian
Lithuanian
Russian
Turkish
Slovak
Slovenian
Croatian
Greek
Estonian
Measuring principle: Colorimetric
Measuring range: LR model: 4 - 3000 µg/L as PO4 [Detection range capable of 4 - 5000 µg/L as PO4]
HR model: 200 - 50000 µg/L as PO4
Mounting: Wall, panel or table
Number of channels: 1, 2, 4; programmable sequence
Operating humidity: 5 - 95 % non-condensing (indoor use only)
Operating temperature range: 5 - 45 °C
Outputs: 4 - 20 mA
Pollution degree: 2/II
Power requirements (Hz): 50/60 Hz
Power requirements (Voltage): 100 - 240 V AC, 24 V DC
Protection class: I
Reagent consumption: 2 L of each reagent every 90 days with 15 minute cycle time
Relays: Four; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC maximum
Repeatability: LR model: ±1%
HR model: ±500 µg/L or ±5% of reading, whichever is greater
Response time: Typically, 9 minutes at 25 °C; changes with temperature
Sample pressure: 0.14 - 6 bar (to preset pressure regulator)
Sample temperature: 5 - 50 °C
Storage conditions: -20 - 60 °C
Weight: 20 kg without reagents and standards, 36.3 kg with reagents